Technical Parameters
Measurement Range: 10~10,000 s/in²·100 mL·1.22 KPa
Pressure Adjustment Range: 0~20 KPa (Other pressures can be customized)
Resolution: 0.1 Pa
Sample Size: Φ12 mm
Permeation Area: 0.025 square inch (0.16 cm²)
Number of Test Samples: 3 or 2 or 1 (piece)
Test Gas: O2, N2, CO2, and other gases with a purity of 99.9% (dry gas, gas source provided by the user)
Gas Source Pressure: 0.4 MPa~0.6 MPa
Connection Size: Φ4 mm polyurethane tube
Power Supply: 220 VAC 50 Hz / 120 VAC 60 Hz
Dimensions (Main Unit): 645 mm (L) × 500 mm (W) × 468 mm (H)
Net Weight: 60 kg
Application Range
The BTY-B2P is used for testing the gas permeability and barrier properties of various materials such as battery separators, breathable membranes, and related polymer products.
Detail
Detail